LC fundas – Low backpressure.

As opposed to high back-pressure, it is relatively simple to figure out why the back-pressure on your HPLC system is below normal.

By “normal” backpressure, one means that you know what the usual backpressure for a given LC column under given operating conditions is.

Since we already know that the usual backpressure on a C18, 150mm, 4.6mm ID, 5ยต column, for AcN:water:;70:30 at 1ml/min at 25 deg C is about 700 to 1000 psi, a backpressure lower than 600 psi requires your attention.

The first suspect is a trapped air-bubble inside the system, especially in the pump piston head. An air-bubble will cause the back-pressure to drop and fluctuate.

Disconnect the column and purge out your HPLC thoroughly. If the backpressure is still lower than normal, it means there’s a leak somewhere.

High-pressure sites, i.e., any fittings and seals before the column are most likely to leak. Check-valves, piston seals and injector fittings, in that order, are what you should be looking at.

When you locate the leaking fitting, do not overtighten it. You can cheat a bit, and use a small piece of teflon tape around the fitting. That should take care of it.

If the threads on the fitting are damaged, you have to replace it. Wherever possible, I use Fingertight fittings, and I’d suggest you try them out too.

Cheers … SKSrinivas.

Advertisements

Leave a Reply

Fill in your details below or click an icon to log in:

WordPress.com Logo

You are commenting using your WordPress.com account. Log Out /  Change )

Google+ photo

You are commenting using your Google+ account. Log Out /  Change )

Twitter picture

You are commenting using your Twitter account. Log Out /  Change )

Facebook photo

You are commenting using your Facebook account. Log Out /  Change )

w

Connecting to %s